Test Searches require one of the following browsers: Google Chrome, Safari, or Firefox. This page WILL NOT WORK with Internet Explorer versions 8 or older.

Alphabetical Index


CHLORIDE, body fluid (8000100620)
Test Mnemonic:

CL BF

Specimen Requirements:
Collection:

10 mL plastic, conical, capped tube. (Min: 1mL of fluid).

Storage/Transport:

Delivered to Sample Management within 4 hours of collection.

Stability:

Room temperature for ≤7 days; Refrigerated for ≤4 weeks; Frozen for indefinite.  

Causes for Rejection:

Specimen too viscous for sampling, incomplete and/or incorrect sample identification, improper storage/transport.                                  

Specimen Preparation:

Centrifuge the specimen prior to analysis to remove possible cellular and bacterial contaminants.

Methodology:

Potentiometric

Performed:

Clinical Chemistry

Turnaround Time:

Routine: 8 hours; ASAP: 4 hours; STAT: 1 hour

Reference Range:

By report (reports may vary based on instrumentation, patient age and sex).

CPT 4 Code:

82438

Note:

Indicate the source of fluid on the request form.

Effective Date: 03/06/2013
Reviewed By/Date: Okorodudu, Anthony - 02/17/2019
Approved By/Date: Okorodudu, Anthony - 12/20/2022
When ordering tests for which Medicare or Medicaid reimbursement will be sought, physicians should only order tests that are medically necessary for the diagnosis or treatment of the patient. Components of the organ or disease panels may be ordered individually. The diagnostic information must substantiate all tests ordered and must be in the form of an ICD-10 code or its verbal equivalent.